LCR Meter – 104A

Model AB-104A is a Micro –Controller based precision LCR meter. With basic accuracy of 0.1 %.it offers various test frequencies ,can display two selected parameters simultaneously, has internal Bias for electrolytic capacitors. Auto mode gives auto ranging & automatic component identification. Four wire measurement minimizes error in reading. RS232 Communication & Binning facility upto 8 bins possible. Loading of pre-setted parameters also possible to save time.

Reading can be Hold for documentation. SER & PAR modes are available.

Instruments are calibrated with our calibration & service facility center with standards having traceability to National standards.

Optional accessories are Kelvin clip, SMD tweezer,capacitor insertion jig.

LCR Meter - 104A
  • Measures any Two parameters of R,L,C , Q & D simultaneously
  • Basic Accuracy 0.1% of reading
  • Full Auto Function & Auto Ranging/User defined functions.
  • User Prompt facility for best accuracy
  • Internal 2 Volt DC bias for Electrolytic capacitor
  • D with 0.0001 resolution
  • Selectable 100Hz(120Hz**) / 1 KHz/10KHz/100KHz measurement frequency.
  • Two LCD 16X2 display(one with big font)
  • Series or parallel measurement modes.
  • Four terminal measurements to increase accuracy
  • Calibration has tracebility to National standards.
  • RS232 communication with various optional software.
  • Limits comparator & Binning facility

Model AB-104A is a Micro –Controller based precision LCR meter. With basic accuracy of 0.1 %.it offers various test frequencies ,can display two selected parameters simultaneously, has internal Bias for electrolytic capacitors. Auto mode gives auto ranging & automatic component identification. Four wire measurement minimizes error in reading. RS232 Communication & Binning facility upto 8 bins possible. Loading of pre-setted parameters also possible to save time.

Reading can be Hold for documentation. SER & PAR modes are available.

Instruments are calibrated with our calibration & service facility center with standards having traceability to National standards.

Optional accessories are Kelvin clip, SMD tweezer,capacitor insertion jig.

Technical Specification

Variables measured

: R, L, C , Q & D

Modes (Equivalent circuit)

: Series & Parallel

Measurement Frequency

: User settable 100Hz (120Hz**)or 1Khzor 10KHz or 100KHz

Accuracy of measurement frequency

: ± 0.01 % of Nominal

Maximum voltage across component

: 0.3Vrms

Measurement update rate

: 2 per second

Maximum Time for valid reading after user connecting component

: 1 second upto 10KHz & 3 second for 100KHz.

Display

: Two LCD 16X2 display(one with big font)

Connection to component under test

: 4 terminals BNC(F)

Trim

:Values can be trimmed as per settings.

Limits comparator

:User defined nominal & HI & LO settings, LCR will display the result either PASS,FAIL HI, FAIL LO

Basic accuracy :± 0.1% of Reading ± 1 digit

Parameter

Measurement Ranges

Ultimate resolution

R

0.001 Ω to 999 MΩ

0.1m Ω

L

0. 001μH to 9900 H

0. 001μH

C

0.001 pF to 9900μF

0. 001pF

Q

0.0001 to 999

0.0001

D

0.0001 to 999

0.0001

Conditions for basic accuracy :

Measurement Frequency 100Hz 1KHz 10 KHz 100KHz
Basic Accuracy ±0.15% ±0.1% ±0.15% ±0.25%
Range of Inductance (Q>10) 4mH- 2000H 400μH-200H 40μH-10H 10μH-10mH
Range of Capacitance (Q>10) 4nF-2000μF 400pF-200μF 40pF –10μF 40pF-0.1μF
Range of resistance (Q<0.1) 2Ω - 1MΩ 2Ω – 500KΩ 2Ω –100KΩ 2Ω – 50KΩ
Q or D 0.25-4 0.25-4 0.25-4 0.25-4
** 120Hz is factory set option for 100Hz

For 0.01<99,extreme case accuracy is 3.5% ± 1 digit.

Input protection

:The input is protected against connection of capacitors of up to 10mF charged to not more than 50V

Operating temperature range : 0 °C to 50 °C
Power Supply : 230VAC ± 10%, 50-60 Hz
Fuse : 20 mm, 250mA (Glass)
Size : 90mm(H),395mm(W),305mm(D)
Weight : Standard 6 Kgs approx.

Optional Accessories : a) Kelvin clip b)SMD Tweezer c)Capacitor Insertion Jig.